Hill-Rom Newborn bassinet; model P247
FDA device recall Z-1027-03 · posted 2003-07-18 · Terminated
Recall details
- Recalling firm
- Hill-Rom, Inc.
- Reason for recall
- The caster/wheel may come off of the bassinet, causing the bassinet to tip.
- Action taken
- Customer notification letters dated June 20, 2003 were sent to each customer. Customers are asked to verify the security of the casters to the base and to visually check for loose or missing caster screws, to remove any bassinets from service until repaired, and that Hill-Rom will visit each customer to replace the caster screws on all units.
- Root cause
- Other
- Status
- Terminated
- Quantity affected
- 1,501
- Distribution
- United States, Brazil, Canada, China, Italy, Japan, Kuwait Netherlands, Qatar, United Kingdom
- Date initiated
- 2003-06-20
- Date posted
- 2003-07-18
- Date terminated
- 2003-12-19
- Location
- Batesville, IN