Atlas FDA

Hill-Rom brand Century+ series bed

FDA device recall Z-0903-03 · posted 2003-05-29 · Terminated

Recall details

Recalling firm
Hill-Rom, Inc.
Reason for recall
Possible electrical shock hazard, as the power cord grounding pin may break off or become detached.
Action taken
Letters dated April 24, 2003 were sent to each consignee instructing them to inspect the beds for power cords with non-flared ground pins, to cut plug off any such cords when found, and replace these plugs with new plugs with the flared ground pin. [The type of plug on the unit can be seen by the naked eye, as the plug is transparent plastic molding.]
Root cause
Other
Status
Terminated
Product code
KNC
Distribution
United States, Bahamas, Bermuda, Canada, Colombia, Costa Rica, Dominican Republic, Guam, Japan, Mexico, Panama, Saudi Arabia, South Korea, Taiwan and Venezuela
Date initiated
2003-04-24
Date posted
2003-05-29
Date terminated
2003-09-25
Location
Batesville, IN