Hill-Rom brand Century+ bed; model P1400.

FDA device recall Z-0338-04 · posted 2004-01-10 · Terminated

Recall details

Recalling firm
Hill-Rom, Inc.
Reason for recall
If fluid is spilled on the siderail or the bed control areas, the bed may change positions without user input.
Action taken
A recall letter dated December 26, 2003 was sent to each customer. Customers are asked to follow instructions in the user manual, if there is a fluid spill onto a unit, until Hill-Rom can inspect each unit to determine whether it is affected by the problem.
Root cause
Other
Status
Terminated
Product code
FNL
Quantity affected
222
Distribution
United States, Canada , Panama and Taiwan.
Date initiated
2003-12-26
Date posted
2004-01-10
Date terminated
2004-05-13
Location
Batesville, IN