Hill-Rom Affinity Three Birthing Bed, Model P3700.

FDA device recall Z-1503-06 · posted 2006-09-14 · Terminated

Recall details

Recalling firm
Hill-Rom, Inc.
Reason for recall
The cable on the auxiliary outlet may become pinched, which may result in an electrical short with melting of the plastic transformer and the emission of smoke.
Action taken
Customers were notified via letter dated 8/18/06 and instructed to either correct the problem themselves with the instructions provided or to contact the firm requesting a visit from the recalling firm to correct their beds.
Root cause
Other
Status
Terminated
Product code
KNC
Quantity affected
9,324
Distribution
Nationwide, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, France, French Polynesia, Germany, Guam, Hong Kong, Iceland, India, Indonesia
Date initiated
2006-08-18
Date posted
2006-09-14
Date terminated
2009-07-14
Location
Batesville, IN