High voltage generator XTP-8100G and XTPG-8100G for the Infinix-i series interventional angiography system.

FDA device recall Z-0198-06 · posted 2005-11-24 · Terminated

Recall details

Recalling firm
Toshiba American Med Sys Inc
Reason for recall
It was found during a study, arcing might occur in the transformer of the high voltage generator, XTP-8100G and XTBP-8100G displaying an error message and interrupting fluoroscopy.
Action taken
Consignees notification that a local TAMS service representative will schedule the implementation of the replacement transformer.
Root cause
Other
Status
Terminated
Product code
JAA
Quantity affected
30
Distribution
CA, CO, KS, IL, LA, IN, ME, MI, NC, NJ, NY, OH, PA, TX
Date initiated
2005-09-21
Date posted
2005-11-24
Date terminated
2007-07-12
Location
Tustin, CA