Heparin Lock Flush Solution 100 Unit/mL, USP 3 ml, 4 ml, 5mL, 6 ml fill in a 10 ml Flush Syringe ZR Excelsior Medical Ne

FDA device recall Z-1707-2012 · posted 2012-06-05 · Terminated

Recall details

Recalling firm
Excelsior Medical Corp
Reason for recall
Some lots of Heparin Lock Flush, USP Syringes (3 mL fill in 10 ml syringe, 4 mL fill in 10 ml syringe, 5 mL fill in 10 ml and 6 ml in 10 ml syringe), failed Anti-Factor IIa assay for potency.
Action taken
Excelsior Medical sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 4, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue using the affected product and to complete and return an attached Customer Reply Form via fax to (866) 807-3714. Contact Excelsior Medical Customer Service at (800) 487-4276 for questions regarding this recall. A second letter which included additional lots was sent out on 5/25/2012 with the date of 5/24/2012. A return response form was also included.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
NZW
Quantity affected
34,230,500 units (596 lots)
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Brazil, Canada, and Colombia.
Date initiated
2012-05-07
Date posted
2012-06-05
Date terminated
2013-09-12
Location
Neptune, NJ