Hemotherm Dual Reservoir Cooler/Heater System Models 400CE and 400MR Product Usage: A dual reservoir, water re-circulati

FDA device recall Z-0629-2018 · posted 2018-02-13 · Terminated

Recall details

Recalling firm
Cincinnati Sub-Zero Products LLC
Reason for recall
Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.
Action taken
Between November 27, 2017 - December 8, 2017, Cincinnati Sub-Zero Products LLC, distributed Urgent Medical Device Correction notices to their customers via Email and USPS. This Medical Device Correction has been initiated to provide a labeling update to the cleaning & disinfection procedures. The update addresses the disinfection process, cleaning of reusable accessories, as well as clarifying the water type that should be used with the Hemotherm. Customers are encouraged to do the following: 1) Discard your current Hemotherm Operation & Technical Manual and/or Operation Manual. 2) Replace your current Hemotherm Operation & Technical Manual and/or Operation Manual with the enclosed updated Manual(s). For your convenience, all updates have been highlighted in the manuals. " For Hemotherm 400CE users, utilize the P/N 56418-R Operation & Technical Manual and P/N 56075-G Operation Manuals. " For Hemotherm 400MR users, utilize the P/N 56234-S Operation Manual. All customers are encouraged to complete and return the enclosed response form by fax or email to FA2016-009@genthermcsz.com; Fax: 513-772-9119. Customers with questions or concerns, please contact the Field Action Coordinator: Kathy DeSmidt, E-mail: FA2016-009@genthermcsz.com, Phone: 513-772-8810 or 800-989-7373.
Root cause
Device Design
Status
Terminated
Product code
DWC
Quantity affected
1355 units
Distribution
US Nationwide Distribution
Date initiated
2017-11-27
Date posted
2018-02-13
Date terminated
2018-07-02
Location
Cincinnati, OH