Hemotherm 400CE, Catalog Numbers 86022 & 86023 The Hemotherm Dual Reservoir Cooler/Heater is used to lower, maintain, or
FDA device recall Z-0063-2019 · posted 2018-10-03 · Terminated
Recall details
- Recalling firm
- Cincinnati Sub-Zero Products LLC, a Gentherm Company
- Reason for recall
- Potential for membrane (control panel) switch errors leading to an inability of the device to change control settings or an unintended change to control settings.
- Action taken
- On September 4, 2018, the firm, CSZ, issued an "Urgent Medical Device Correction" notices to customers overnight via courier service. Customers were advised to take the following actions: - Immediately examine your inventory and quarantine product subject to the field action. - If you choose to use the product prior to the correction, please consider having a backup device available. - In addition, if you have further distributed this product, please identify your customers and notify them at once of this medical device correction. Your notification to your customers may be enhanced by including a copy of the Urgent Medical Device Correction letter. - Contact CSZ Technical Support (1-888-437-5608) for a replacement kit: Reference FA2017-015 for the Hemotherm control panel kit or you may contact CSZ Technical Support to request replacement by a CSZ representative. - Once the replacement kit is received follow Hemotherm Control Panel Replacement Instructions provided in the kit. - Complete and return the response form provided to acknowledge receipt of the notification and to inform CSZ that customers have performed the requested actions via fax to (513) 772-9119 or via email to FA2017-015@genthermcsz.com. If you have any questions, call Kathy DeSmidt @ 513-772-8810.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- DWC
- Quantity affected
- 270 units
- Distribution
- US Nationwide Distribution
- Date initiated
- 2018-09-04
- Date posted
- 2018-10-03
- Date terminated
- 2019-05-23
- Location
- Cincinnati, OH