Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatogra

FDA device recall Z-1997-2013 · posted 2013-08-22 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
Beckman Coulter has received a customer complaint that suggests trending of positive results has increased from approximately 4% to 1 0%. As a result of our investigation, Beckman Coulter has confirmed an increase in sensitivity at the borderline positive level. " This testing also demonstrated that the issue does not affect results reported as negative.
Action taken
Beckman Coulter sent an Urgent Product Correction letter dated July 9, 2013, to all affected customers. The letter was accompanied with a cover letter and response form, dated July 9, 2013. Customers were informed of the product under recall with product description, actions to take, contact information, the issue, impact, and resolution. Beckman Coulter conducted a risk assessment for this issue. Per the risk documentation, there is only a remote probability that use of the product will cause any adverse health consequences,and the product is not likely to lead to serious adverse health consequences. If you have any questions please call (714) 961-3777.
Root cause
Under Investigation by firm
Status
Terminated
Product code
KHE
Quantity affected
68,116 units
Distribution
Worldwide Distribution - USA and the country of Canada.
Date initiated
2013-07-17
Date posted
2013-08-22
Date terminated
2014-10-24
Location
Brea, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
HEMOCCULT ICT (K080812)Beckman Coulter, Inc.2008-06-25