HeartWare Ventricular Assist System (HeartWare Controller) Product Usage: For use as a bridge-to-cardiac transplantation
FDA device recall Z-1131-2015 · posted 2015-02-25 · Terminated
Recall details
- Recalling firm
- HeartWare Inc
- Reason for recall
- The affected clinical trial Controllers exhibit a higher susceptibility to ESD than newer commercial Controllers.
- Action taken
- HeartWare sent an URGENT MEDICAL DEVICE RECALL dated January 16, 2015 asking clinicians to quarantine the affected controllers and complete and return the acknowledgement form. HeartWare also asks clinicians to identify patients with affected controllers and review the applicable risks with the patient and as soon as reasonably possible if medically advisable, exchange the affected controller under medical supervision with a new controller. HeartWare also asks that for each patient using an affected controller as their backup to contact the patient and arrange to have the backup controller replaced with a new controller. Lastly, HeartWare asks the firm to complete the patient tracking form and return all quarantined controllers to HeartWare along with the completion form. Should you have any questions or concerns, please contact your local HeartWare representative. Our 24-Hour Clinical Support personnel are also available at 1-888-494-6365.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- DSQ
- Quantity affected
- 4,845
- Distribution
- Worldwide Distribution - US Nationwide in the states of: AL, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NY, OH, OR, PA, TX, UT, VA, WA and District of Columbia and countries of: Canad
- Date initiated
- 2015-01-16
- Date posted
- 2015-02-25
- Date terminated
- 2020-02-05
- Location
- Miami Lakes, FL