HeartStart MRx monitor/defibrillator Models M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6
FDA device recall Z-0219-2016 · posted 2015-11-05 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems, Inc.
- Reason for recall
- MRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed
- Action taken
- Philips Healthcare issued on 12/23/14 The Urgent Medical Device Correction notification/Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. A software upgrade will be provided free of charge for all units affected by one or more of these issues. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the upgrade. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: You can continue to use your MRx prior to receiving the software upgrade. If the MRx reboots during clinical use, continue to treat the patient according to your organizations protocol, reinitiating therapy if necessary.
- Root cause
- Software design
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 75,693 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of AFGHANISTAN, ALBANIA, ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOWINA, BRAZI
- Date initiated
- 2014-12-23
- Date posted
- 2015-11-05
- Date terminated
- 2020-09-09
- Location
- Andover, MA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| HEARTSTART MRX, MODELS M3535A OR M3536A (K063375) | Philips Medical Systems | 2007-01-11 |
| HEARTSTART MRX MONITOR/DEFIBRILLATOR (K062233) | Philips Medical Systems North America, Inc. | 2006-11-22 |
| HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS (K061707) | Philips Medical Systems | 2006-08-11 |
| HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A (K051134) | Philips Medical Systems | 2005-09-08 |
| HEARTSTART MRX, MODEL M3535A (K040404) | Philips Medical Systems, Inc. | 2004-05-04 |
| HEARTSTART MRX, MODEL M3535A (K031187) | Philips Medical Systems | 2003-08-27 |