HeartStart MRx Defibrillator/Monitor, M3535A/M3536A Is used for the termination of ventricular tachycardia and ventricul
FDA device recall Z-1390-2009 · posted 2009-06-02 · Terminated
Recall details
- Recalling firm
- Philips Healthcare Inc.
- Reason for recall
- Spontaneous turn-on which could deplete the battery, rendering the device unusable until power is restored. There is also the possibility of a failure mode in which the device fails to respond to user initiated turn-on, rendering it unusable for monitoring and therapy.
- Action taken
- Philips notified customers via Urgent Medical Device Recall Letter & Notice dated March 2009. Customers are asked to follow the "Actions to be taken by customer/user section of the Urgent Medical Device Recall Notice. In this section, customers are given the following instructions: "During the interim period, as you await the upgrade for your device, you may continue to use the Heartstart MRx. However, to reduce the risk that a spontaneous turn-on could deplete the battery, it is advised that you keep: a charged spare battery with the Heartstart MRx AC power applied, if possible, when the device is not in use You should also consider identifying a readily available backup defibrillator to use in the unlikely event that the device fails to turn on. Devices that exhibit a spontaneous turn-on or failure to respond to user initiated turn on should be removed from service as soon as possible." Philips will replace the therapy switch in affected units free of charge per Mandatory Field Change Order FC086100077. All customers will be contacted by a Philips Healthcare representative to arrange for replacement of the therapy switch in the affected devices.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 12,385 units US; 9,791 units OUS
- Distribution
- Worldwide Distribution -- USA, Canada, Afghanistan, Argentina, Aruba, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Czech R
- Date initiated
- 2009-04-10
- Date posted
- 2009-06-02
- Date terminated
- 2016-12-08
- Location
- Andover, MA