HeartStart FR2+ automated external defibrillator (AED) model: M3861: PHILIPS brand; text prompts only on screen (NO ECG)
FDA device recall Z-0472-2011 · posted 2010-11-26 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems
- Reason for recall
- Failure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
- Action taken
- Philips Medical Systems issued an Urgent Medical Device Recall letter dated May 28, 2010 identifying the affected device and actions to be taken. Customers were instructed to review their inventory for affected devices. If affected devices are found, the customer should contact Philips to arrange for a replacement device. The customer should also run a battery insertion test. If the device passes the test, the AED may remain in service until a replacement is received. For further information, contact Philips at 1 800 263-3342, option 6.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 13,355 units
- Distribution
- Worldwide Distribution: Throughout the USA, and to the following countries: ALBANIA, ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BERMUDA, BOSNIA AND HERZEGOVINA, BRAZIL, CANADA, CHILE, CHI
- Date initiated
- 2010-05-28
- Date posted
- 2010-11-26
- Date terminated
- 2013-09-13
- Location
- Seattle, WA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| PHILIPS HEARTSTART FR2+ AUTOMATED EXTERNAL DEFIBRILLATOR, MODELS M3860A, M3861A, M3840A, M3841A (K051632) | Philips Medical Systems | 2005-10-05 |
| HEARTSTREAMFR2 AED WITH M3848A AND M3849A, MODELS M3860A, M3861, M3840A, M3841A (K014157) | Philips Medical Systems | 2002-01-17 |
| HEARTSTREAM/HEARTSTART FR2 AED ECG CABLE, MODELS M3860A, M3861A, M3840A, M3841A (K013425) | Philips Medical Systems | 2002-01-14 |
| HEARTSTREAM ATTENUATED DEFIBRILLATION PADS, MODEL M3870A (K003819) | Agilent Technologies, Inc. | 2001-05-02 |
| HEARTSTREAM FR2 AED, MODEL M3860A,M3861A (K003565) | Agilent Technologies, Inc. | 2000-12-20 |