HeartStart FR2+ AED with Text Display (No ECG), Model M3861, Philips brand, no configurable manual charge. The label of
FDA device recall Z-0484-2011 · posted 2010-11-29 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems
- Reason for recall
- A flash memory component in the HeartStart FR2+ and Automated External Defibrillator was improperly relabeled by a third party. Therefore the component could not be identified by its source, quality, and date of manufacture.
- Action taken
- On 10/5/10, Philips began sending out the Voluntary Medical Device Removal notification letter to their consignees. This letter informs the consignees that Philips Healthcare is conducting a voluntary product removal of a limited number of HeartStartFR2+ automated external defibrillators (AEDs), model numbers M3860A, M3861A, M3849A, and M3841A, manufactured by Philips and shipped between March 2007 and August 2010. The owners of affected AEDs are advised to run a battery insertion test. If it passes, the AED should remain in service until its replacement is received. Philips will be replacing affected units. The consignees can visit the firm's website www.philips.com/FR2AEDAction or contact Philips at 1-800-263-3342 for any information or support concerning the issue.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 11, 158 units
- Distribution
- Devices were distributed to hospitals, medical facilities, and distributors through out the US and to foreign consignees. Foreign consignees including countries: Algeria, Argentina, Australia, Austria
- Date initiated
- 2010-09-22
- Date posted
- 2010-11-29
- Date terminated
- 2013-09-05
- Location
- Seattle, WA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| HEARTSTREAM/HEARTSTART FR2 AED ECG CABLE, MODELS M3860A, M3861A, M3840A, M3841A (K013425) | Philips Medical Systems | 2002-01-14 |
| HEARTSTREAM FR2 AED, MODEL M3860A,M3861A (K003565) | Agilent Technologies, Inc. | 2000-12-20 |