Atlas FDA

HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-

FDA device recall Z-1141-2024 · posted 2024-02-15 · Open, Classified

Recall details

Recalling firm
HeartSine Technologies Ltd
Reason for recall
Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.
Action taken
On 2/14/24, correction notices were mailed to customers who were asked to do the following: 1) Maintain awareness of this communication internally until the required action has been completed within your facility. 2) Inform the recalling firm if any of the subject devices were distributed to other organizations. The firm will work with you to ensure recipients are notified appropriately. 3) Complete and return the acknowledgment form to RSRecall@Stryker.com Upon receipt of the acknowledgment form, the firm will provide instructions on next steps including device replacement. If you have any questions or concerns, please contact the firm's Customer Support at +44 28 9093 9400, 9:00 A.M. to 5:00 P.M. (GMT), Monday - Friday or by email at heartsinesupport@stryker.com.
Root cause
Process control
Status
Open, Classified
Product code
MKJ
Quantity affected
12
Distribution
Worldwide - US Nationwide distribution in the states of TX, CA, IL, CO, FL, NC, MI and the countries of Netherlands, Japan, New Zealand.
Date initiated
2024-02-14
Date posted
2024-02-15
Location
Belfast, United Kingdom

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.