Heartsine Samaritan AED Defibrillators - Models: SAM 001, SAM 002, SAM 003. Serial numbers within range of 1270 - 2324 (
FDA device recall Z-0641-05 · posted 2005-03-23 · Terminated
Recall details
- Recalling firm
- Heartsine Technologies Inc
- Reason for recall
- The affected Automatic External Defibrillators (AED) annouce a fault warning and shut themselves off before a shock can be delivered to the patient. This is reportedly due to slow capacitor charging rates which are interpreted by the AED software as a fault condition.
- Action taken
- Recall was initiated by telephone 2/14/2005 with follow-up fax, email or postal mailing using one of several customized letters for distributors or customers. Customers will be sent an upgrade kit by overnite mail and forms to confirm upgrade.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 672
- Distribution
- Nationwide to distributors and user facilities. Also to Argentina, Austria, Belgium, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Portugal, Puerto Rico, Sin
- Date initiated
- 2005-02-14
- Date posted
- 2005-03-23
- Date terminated
- 2011-06-08
- Location
- San Clemente, CA