HeartSine SAM 350P, HeartSine SAM 360P, and HeartSine SAM 450P The HeartSine samaritan PAD (Public Access Defibrillator)
FDA device recall Z-2174-2025 · posted 2025-07-25 · Open, Classified
Recall details
- Recalling firm
- HeartSine Technologies Ltd
- Reason for recall
- Due to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks)
- Action taken
- On 06/30/2025, the firm sent via USPS First Class an "URGENT: MEDICAL device recall" Letter to customers informing them Stryker has determined that for the HeartSine samaritan PAD, a manufacturing process issue related to a circuit board component may impair the device's ability to function or cause failure. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. The device becomes inoperable after the failure occurs. Customer are instructed to: 1. Identify impacted devices by verifying if their device serial numbers are included in the list at the link https://www.stryker.com/us/en/emergency-care/product-notice/heartsine/index.html 2. Submit customer response electronically via the QR code scan or link https://bit.ly/fa318-heartsine, or response via email to: RSRecall@stryker.com and include the following Subject: FA 318 HeartSine T3 - Response from <<>> Email Body: -Customer ID, Customer Name, title and email address -The number of devices impacted by model, and serial numbers -Provide if any devices have been further distributed 3. Until a replacement is available, Stryker recommends keeping HeartSine device in service if there is no alternative public access defibrillator. For questions or concerns, contact Stryker Customer at +1 800-787-9537, option 2 from 8:00 AM to 7:00 PM (Eastern Time) Monday thru Friday or contact Technical Support at heartsinesupport@stryker.com
- Root cause
- Process change control
- Status
- Open, Classified
- Product code
- MKJ
- Quantity affected
- 190,516 units
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM,
- Date initiated
- 2025-06-30
- Date posted
- 2025-07-25
- Location
- Belfast, United Kingdom
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.