HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with Hear

FDA device recall Z-1570-2025 · posted 2025-04-17 · Open, Classified

Recall details

Recalling firm
Thoratec LLC
Reason for recall
Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.
Action taken
On 3/13/2025, recall notices were sent to customers who were asked to do the following: If patients are/ and are not able to be identified using the serial number list, send those patients the patient letter. If the MPU is not listed, this device is not affected. For units in use that have experienced MPU power issues: If there is an MPU power failure, transfer the patient from the MPU to the 14V rechargeable batteries within 15 minutes (see IFU). The Backup Battery in the System Controller will temporarily power the pump during a power source switch for up to 15 minutes. MPU will continue to echo the controller alarms as long as the internal AA batteries in the MPU do not need replacement. Do not continue to use the MPU and contact firm's Technical Support at 1-800-456-1477 (8 AM EST to 7 PM EST) for return and replacement. For units in use but not experiencing MPU power issues: Provide patient letter and educate about the issue. Ensure the patient's 14V rechargeables are ready and available for use. Prompt patient to replace AA batteries inside the MPU immediately if the Yellow Mobile Power Unit Battery Indicator alarm is active. Firm will replace these with new units starting June 2025. Return/replace units not distributed, not in use, on shelves, or that are unopened. Contact Customer Support at 1-612-444-3013, Option 2 (9 AM EST to 6:30 PM EST). Complete/return the acknowledgement form via email to MCSHMNotices@abbott.com Distribute this notice to those who need to be aware within your institution and to any organization to which these devices may have been transferred. Device lookup tool with the impacted MPU devices, available by March 24th, 2025: https://www.cardiovascular.abbott/us/en/hcp/products/heart-failure/left-ventricular-assistdevices/ heartmate-3/indications-safety-warnings/advisories.html by March 24th, 2025. If you have questions, contact Technical Support.
Root cause
Process control
Status
Open, Classified
Product code
DSQ
Quantity affected
1661
Distribution
Worldwide - US Nationwide distribution in the states of WA, NE, VA, PA, DC, KY, CA, IL, WI, TX, MA, AR, MS, MI, OH, FL, NC, TN, LA, NJ, UT, NY, DE, CT, GA, IN, AZ, OK, MO, OR, MN, SC, NM, MD, IA, KS,
Date initiated
2025-03-13
Date posted
2025-04-17
Location
Pleasanton, CA