HeartMate II LVAS with Pocket Controller; 106762 - HMII POCKET CONTROLLER, NA; UDI: 00813024011286 Intended to provide h
FDA device recall Z-2053-2017 · posted 2017-05-24 · Terminated
Recall details
- Recalling firm
- Thoratec Corporation
- Reason for recall
- St. Jude Medical is providing all users of their HeartMate II LVAS with Pocket Controller with new software and updates to hardware to make the exchange to a backup controller easier. This is related to recall Z-1227/1230-2014.
- Action taken
- Urgent Medical Device Correction letters, dated March 29, 2017, were sent to customers. The letters provided information on what the current correction will involve to ensure patients are successfully able to exchange their pocket controller in an emergency situation. Patients will receive the updates at scheduled appointment. Questions regarding the notice, should be directed to Abbott HeartLine at 800-456-1477; or, your Abbott Sales Representative.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- DSQ
- Quantity affected
- 24,077 active units in US
- Distribution
- Worldwide Distribution -- US and Austria, Belgium, Brazil, Canada, Cayman Islands, Chile, Columbia, Croatia, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Greece, Hong Kong, Hungary, India
- Date initiated
- 2017-03-30
- Date posted
- 2017-05-24
- Date terminated
- 2019-03-20
- Location
- Pleasanton, CA