Atlas FDA

Harvard 2 Syringe Pump manufactured for Rita Medical (OEM) Model Number 2005-001

FDA device recall Z-0433-05 · posted 2005-01-26 · Terminated

Recall details

Recalling firm
Harvard Clinical Technology
Reason for recall
Potential pump motor/encoder assembly failure may cause over infusing medication in patients
Action taken
Harvard Clinical Technology notified consignees by letter on 11/3/04. Harvard will contact consignees by telephone to schedule a return. The motors inside the recalled products will be replaced at Harvard Clinical. Harvard will provide Loaner pumps upon request during the repair process.
Root cause
Other
Status
Terminated
Product code
MEA
Quantity affected
204 units
Distribution
AK, KY, TX, MA, NY, CA, GA, OK, IL, CT, LA, FL, MD,NC, NJ, VA, MI, WI, IA, WV,OH,
Date initiated
2004-11-03
Date posted
2005-01-26
Date terminated
2012-05-07
Location
Natick, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
HARVARD 1 SYRINGE PUMP/HARVARD 1 WITH PCA SYRINGE PUMP (K033054)Harvard Clinical Technology2003-12-16