Atlas FDA

Harleco May and Grunwald's Stain; 0.17% w/v; Methanol Solution UN1230; EMD Chemicals Inc. An Affilifate of Merck KGaA Da

FDA device recall Z-0133-2008 · posted 2007-11-03 · Terminated

Recall details

Recalling firm
EMD Chemicals Inc
Reason for recall
Out of Spec -- The Batch (Lot) was incorrectly released after being rejected for out of specification of high azure and eosin absorbance at the end of the run and particulate matter at the end of the run.
Action taken
EMD Chemicals via fax and mailed an Important Product Recall letter dated 7/30/07. The letter notified end use consignee that after conducting an investigation it has determined that Harleco May and Grunwald's Stain; 0.17% w/v Item #660/75 % 660/85 has unacceptable levels of azure and eosin absorbance and undissolved dye matter. Users were requested to discontinue shipping and return product. Wholesalers were advised this is a Customer Level Product Recall and requested to sub recall product sold after 7/5/07.
Root cause
Employee error
Status
Terminated
Product code
KQC
Quantity affected
26 units (1 liter), 1 unit (4 liter)
Distribution
Product was released to 2 distributors; one in the US, one in Canada
Date initiated
2007-07-30
Date posted
2007-11-03
Date terminated
2007-12-26
Location
Gibbstown, NJ