Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator

FDA device recall Z-2526-2023 · posted 2023-09-20 · Open, Classified

Recall details

Recalling firm
Hamilton Medical, Inc.
Reason for recall
Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
Action taken
On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
CBK
Quantity affected
76 units
Distribution
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Date initiated
2023-07-26
Date posted
2023-09-20
Location
Reno, NV