Hamilton-MR1 Ventilator: Catalog# 161010 The Hamilton MR1 Ventilator is intended to provide pressure ventilator support

FDA device recall Z-0666-2017 · posted 2016-11-28 · Terminated

Recall details

Recalling firm
Hamilton Medical, Inc.
Reason for recall
Oxygen tubing and the oxygen connector of the Hamilton-MR1 could potentially become loose during the preparation for ventilation.
Action taken
Hamilton Medical sent a letter dated September 6, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Hamilton Medical will update each affected Hamilton-MR1 ventilators by installing a new oxygen mixer mounting plate with the correct screws at no cost to each facility. A Hamilton Medical Field Service Technician will contact customers soon to schedule a convenient time to install the hardware upgrade. Customers may also call 1-800-426-6331, option #2 to be issued an RGA for return to the service center for the upgrade kit to be installed. Customers with questions should call 817-909-0308.
Root cause
Device Design
Status
Terminated
Product code
CBK
Quantity affected
44 units
Distribution
Nationwide Distribution
Date initiated
2016-09-26
Date posted
2016-11-28
Date terminated
2017-03-08
Location
Reno, NV