Hamilton Medical Infant Flow Sensor, single use, (1.88m), Part number 155500. The infant flow sensor is an accessory for

FDA device recall Z-1964-2015 · posted 2015-07-02 · Terminated

Recall details

Recalling firm
Hamilton Medical, Inc.
Reason for recall
An issue has been discovered with the Hamilton Medical Infant Flow Sensor, single use. When a ventilation mode with adaptive volume control (APVcmv/CMV+) is used, the ventilator may adjust the ventilation on the inaccurate flow measurement and decrease the amount of ventilation delivered.
Action taken
Hamilton Medical Inc sent a Medical Device Safety Notices letter dated May 8, 2015 by traceable method to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact Technical Support Manager at 800-426-6331, Ext 215.
Root cause
Device Design
Status
Terminated
Product code
BZO
Quantity affected
154 total, all part numbers
Distribution
US Nationwide Distribution.
Date initiated
2015-06-01
Date posted
2015-07-02
Date terminated
2015-08-12
Location
Reno, NV