HAMILTON-G5, with software versions less than or equal to 2.60 The HAMILTON-G5/S1 ventilators are designed for intensive
FDA device recall Z-1874-2019 · posted 2019-07-11 · Terminated
Recall details
- Recalling firm
- Hamilton Medical AG
- Reason for recall
- New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.
- Action taken
- On 03/25/19, the firm, Hamilton AG, issued Medical Device Field Safety Corrective Action notices to customers via certified mail with receipt response. Customers were advised that if ventilators entered the ambient state, they should provide alternate ventilation immediately. In addition, in order to exit the ambient state, you must turn off ventilator power. The firm has developed a new software version, which reduces the probability of the ventilator entering an ambient state due to the occurrence of an error which leads to the user message "panel connection lost". The firm will contact customers to schedule the software update. The firm requests that consignees distribute this notification to all end users within their organization. They also request that consignees complete and return the acknowledgement form via email: nicole.robinson@hamiltonmedical.net; fax: (775) 856-5621, or mail: Hamilton Medical, Inc., PO Box 30008, Reno, NV 89520. Customers with questions are encouraged to call 1-800-426-6331, option #2. In addition, with the Medical Device Field Safety Corrective Action notices, the firm also provided customers with Medical Device Field Safety Corrective Action notices, dated 02/11/19, sent to OUS customers for Hamilton-G5 and Hamilton-S1 ventilators. Customers were advised that the Swiss firm has developed a new software for the HAMILTON-G5/S1 ventilators. The software version 2.80 reduces the probability of the ventilator entering an ambient state due to the occurrence of an error which leads to the user message "panel connection lost". When the HAMILTON-G5/S1 ventilator enters an ambient state, the following conditions apply: - The HAMILTON-G5/S1 releases both a visual and an acoustical high-priority alarm to alert medical staff. - The inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. - Provide alternative ventilation immediately. - You must turn off ventilator power to exit the Am
- Root cause
- Software design
- Status
- Terminated
- Product code
- CBK
- Quantity affected
- 4338 units
- Distribution
- Worldwide Distribution- US (Nationwide Distribution): MI, WI, OR, MO, TX, NV, PA, SC, HI, MN, ND, NC, AZ, FL, CA, MA, GA, NY, VA, WA, IN, CO, LA, IL, OH, TN, SD, OK, MI, AR, C, AK, WV, NJ, IA, NE, KS,
- Date initiated
- 2019-03-25
- Date posted
- 2019-07-11
- Date terminated
- 2024-06-03
- Location
- Bonaduz, Switzerland
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| HAMILTON-G5 (K131774) | Hamilton Medical AG | 2013-10-09 |
| HAMILTON-G5 (K103803) | Hamilton Medical AG | 2011-11-23 |
| MODIFICATION TO HAMILTON-G5 (K081521) | Hamilton Medical AG | 2008-09-25 |