Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700..

FDA device recall Z-2057-2015 · posted 2015-07-10 · Terminated

Recall details

Recalling firm
Hamilton Medical, Inc.
Reason for recall
Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.
Action taken
*** Please note *** This is not a new recall. An error occurred, and this recall was originally classified as a class I on July 10, 2015. No further action is required. Hamilton Medical AG notified their customers with a letter dated February 19, 2015 and had a fix (SW 2.42) that was available as of February 20th 2015. Hamilton Medical proceeded to send out notification letters the 24th of March 2015 once they compiled the complete and validated list of customers that had the affected software. Customers with questions were instructed to call 800-426-6331, ext.215.
Root cause
Device Design
Status
Terminated
Product code
CBK
Quantity affected
248
Distribution
Nationwide Distribution
Date initiated
2015-03-24
Date posted
2015-07-10
Date terminated
2015-12-23
Location
Reno, NV