Atlas FDA

Gyrus PlasmaKinetic Super Pulse Generator is a micro-computer based radio frequency generator with electrical and softwa

FDA device recall Z-0841-04 · posted 2004-07-20 · Terminated

Recall details

Recalling firm
Gyrus Medical, Inc
Reason for recall
The generator may malfunction when used in close proximity to a monopolar generator.
Action taken
On December 05, 2003, the firm sent recall notices to their affected customers by Certified Mail. The customers were notified that the firm would call them to arrange for pick up and correction of the device.
Root cause
Other
Status
Terminated
Product code
GEI
Quantity affected
14
Distribution
Nationwide throughout the United States.
Date initiated
2003-12-05
Date posted
2004-07-20
Date terminated
2006-12-28
Location
Maple Grove, MN

Monitor recalls programmatically