Gyrus ACMI PKS Seal Open Forceps, model 915005PK, 9-3/4" Straight Jaw. Sterile EO. The device is intended to be used wit

FDA device recall Z-0636-2010 · posted 2010-01-14 · Terminated

Recall details

Recalling firm
Gyrus ACMI Corporation
Reason for recall
Gyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field. The detached shim may be undetected and be left behind in the patient. This circumstance could subsequently cause complications. Further use of this product should cease, affected product should be removed from stockrooms and returned to Gyrus ACMI.
Action taken
An Urgent Medical Device Recall letter, dated 08/15/09 was sent to customers notifying them of the issue. The letter identified affected product, asked that they cease further use, remove affected devices from stockrooms, contact Gyrus ACMI customer service for a return goods authorization and return product to Gyrus ACMI at no charge. A reply form is also asked to be returned to Gyrus ACMI via fax. Questions and requests for a Return Goods Authorization are directed to Gyrus ACMI customer service at 1-888-524-7266.
Root cause
Environmental control
Status
Terminated
Product code
GEI
Quantity affected
93,665 total (63,418 US, 30,247 OUS)
Distribution
US and International distribution. OUS to Include: Australia, Belgium, Bolivia, Bulgaria, Canada, Chile, China, Columbia, Croatia, Czech, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece ,
Date initiated
2009-08-18
Date posted
2010-01-14
Date terminated
2012-03-06