GYC-1000 Green Laser Photocoagulator system The Nidek Green Laser Photocoagulator Model GYC -1000 is indicated for use i

FDA device recall Z-2120-2014 · posted 2014-08-13 · Terminated

Recall details

Recalling firm
Nidek Inc
Reason for recall
Laser Aperture label was not applied to certain GYC-1000 Green Laser Photocoagulators devices in the field as per FDA regulation,
Action taken
Nidek sent an Medical Device Correction Notifications dated May 16, 2014, informing doctors of the lack of the Laser Aperture Label for GYC-1000 Green Laser Photocoagulator System. Letters included a Aperture label and instruction/illustration on where to affix it to the device. A response form is to be completed and returned indicating that they read the correction information and have affixed the label on the GYC-1000 device as instructed. Questions may be directed to Nidek Customer service at 1-800-223-9044 Monday - Friday from 8 am to 5 pm PST. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. You may proceed or continue with implementation of the CAP (subject to the conditions noted above).
Root cause
Radiation Control for Health and Safety Act
Status
Terminated
Product code
GEX
Quantity affected
567 units
Distribution
USA (nationwide) Distribution
Date initiated
2014-05-16
Date posted
2014-08-13
Date terminated
2014-10-16
Location
Fremont, CA