Guidant INSIGNIA I AVT family of pacemakers includes the following: SSI (model 482), VDD (model 882), DDD (model 982), S

FDA device recall Z-0194-05 · posted 2005-11-23 · Terminated

Recall details

Recalling firm
Guidant Corporation
Reason for recall
Another failure mode, which occurs at a low rate, has been identified with the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include; intermittent or permanent loss of pacing output without warning, intermittent or permanent loss of telemetry and appearance of a reset warning message upon interrogation.
Action taken
Important Medical Device Safety Information & Corrective Action , dated 09/22/05 was sent to all affected physicians. This letter explained the failure mode, what devices are affected and gave recommendations. A press release was issued on 09/22/05. UPDATE: Guidant is issuing an Advisory Update to physicians in December 2005. This update provides information relative to the failure mode including the identified cause; FDA's classification; Guidant representatives will retrieve and replace any hospital inventory with product free from "Mode 2" peri-implant failure; and recommended actions.
Root cause
Other
Status
Terminated
Product code
DXY
Quantity affected
14,404
Distribution
Nationwide and Arab Emerites, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Ge
Date initiated
2005-09-22
Date posted
2005-11-23
Date terminated
2007-01-05
Location
Saint Paul, MN