FDA device recall Z-0826-06 · posted 2006-05-02 · Terminated

Recall details

Recalling firm
Guidant Corporation
Reason for recall
77 distributed devices may have a cracked layer of insulation in a flexible hybrid circuit, leading to internal shorting during charging such as during routine capacitor formation or for defibrillation. This would likely render the device non-functional, resulting in no output and/or telemetry communications.
Action taken
Guidant Sales Representatives are retrieving non-implanted devices April 10-12, 2006. If a device is implanted the physician is sent an Urgent Medical Device Safety Information & Corrective Action letter, April 11-14, 2006. This letter describes the issue, states that Guidant is retrieving all non-implanted devices and that Guidant is currently conducting tests on retrieved devices. Physicians will be advised once more information is available.
Root cause
Other
Status
Terminated
Product code
NIK
Quantity affected
20 devices
Distribution
Nationwide. AZ, FL, GA, IL, IN, KY, MA, ME, MI, MN, NC, NE, NY, OH, PA, TN, TX, UT, VA, WI. OUS to include: Australia, Canada, Germany, Italy, Netherlands, Portugal, Slovenia, Spain.
Date initiated
2006-04-07
Date posted
2006-05-02
Date terminated
2007-04-01
Location
Saint Paul, MN