Guidant ASSURE, Model B301, Implantable Cardioverter Defibrillator, ICD. (This model is not available in the USA) Guidan
FDA device recall Z-0049-2008 · posted 2007-11-24 · Terminated
Recall details
- Recalling firm
- Boston Scientific CRM Corp
- Reason for recall
- Atypical Charge Time Behavior: 1)-End of Life (EOL) or ERI is displayed during mid-life (typically 24-48 months), even though battery capacity remains available. 2)- Extended ERI charge time limits: Charge times during mid-life may remain below a normal extension of the ERI charge time . ****** Prophylactic replacement of these devices prior to ERI is not recommended. As stated in the Product
- Action taken
- A Product Update titled "ERI Charge Time Limit Extended During Mid-Life and Mid-Life Display of Replacement Indicators", dated 03/10/2007 was posted to Company's website on 03/13/2007. A CRM Product News letter was sent to physicians beginning 03/16/2007. The CRM Product News Letter gave an overview of the content of the Product Update and stated the web location for finding the complete Product Update. The Product Update identified affected devices, gave an overview of normal charge time behavior at mid-life for specific ICD and CRT-D devices. The letter also described the atypical charge time behavior, projected rate of mid-life display of replacement indicators and Patient Management Considerations. ****** Prophylactic replacement of these devices prior to ERI is not recommended. As stated in the Product Update, "Devices that have triggered charge time-based ERI or EOL during mid-life have several months, and in most cases more than one year of remaining battery voltage and capacity, which allows the devices in this pattern to continue to provide brady and LV pacing and maximum energy shocks. However, if ERI or EOL is triggered, device replacement should be scheduled."*****Recalled units only cover devices implanted prior to July 2005****
- Root cause
- Device Design
- Status
- Terminated
- Product code
- LWP
- Quantity affected
- 180-Recalled units only cover devices implanted prior to July 2005
- Distribution
- Worldwide
- Date initiated
- 2007-03-16
- Date posted
- 2007-11-24
- Date terminated
- 2007-11-23
- Location
- Saint Paul, MN