Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the

FDA device recall Z-2442-2015 · posted 2015-08-21 · Terminated

Recall details

Recalling firm
Medtronic MiniMed Inc.
Reason for recall
Medtronic MiniMed is recalling the Guardian Real-Time Continuous Glucose Monitoring System because it has a language translation error that impacts the Finnish language. Specifically, the Monitor has a Finnish translation error in Predictive Alerts setting: the on-screen Finnish translation for LOW/HIGH is reversed as HIGH/LOW.
Action taken
A Urgent Medical Device recall letter was sent to customers on 7/22/15 to inform them that the Guardian Real-Time Continuous Glucose Monitoring System System has a language translation error that impacts the Finnish language. The letter informs the customers of the problems identified and the actions to be taken. Customers with questions or concerns are instructed to contact Nordic Support Line 0-800-164-064.
Root cause
Software design
Status
Terminated
Product code
MDS
Quantity affected
254 units
Distribution
Distributed in Finland.
Date initiated
2015-07-22
Date posted
2015-08-21
Date terminated
2015-09-18
Location
Northridge, CA