Atlas FDA

Green Spec Fiberoptic Laryngoscope Handle - Stubby/Short; a battery operated fiberoptic laryngoscope handle for use with

FDA device recall Z-1225-06 · posted 2006-07-13 · Terminated

Recall details

Recalling firm
Teleflex Medical
Reason for recall
Teleflex Medical has identified that the product may malfunction, causing the handle to heat up. There is a potential for the heated handle to burn the user.
Action taken
Teleflex Medical sent Urgent Medical Device Recall letters dated 6/15/06 to the direct accounts, informing them of the potential for the handle to heat up, and requesting that they cease use of the affected units and return them to Teleflex Medical for replacement. The accounts were instructed to complete and return by fax the enclosed Acknowledgement & Stock Status Form, indicating the number of units being returned for replacement. Any questions were directed to Linda Todd at 1-800-334-9751, ext. 4951.
Root cause
Component change control
Status
Terminated
Product code
CCW
Quantity affected
1,408 laryngoscopes
Distribution
Nationwide and Canada, Colombia, Costa Rica and Chile.
Date initiated
2006-06-15
Date posted
2006-07-13
Date terminated
2009-08-28
Location
Bannockburn, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
LARYNGOSCOPE BLADES (K883414)Truphatek, Ltd.1988-08-19