Gluteraldehyde Concentration Indicators. MetriTest 1.5% and MetriTest 1.8%, also OmniChek 1.5% and OmniCheck 1.8%

FDA device recall Z-0712-03 · posted 2003-04-09 · Terminated

Recall details

Recalling firm
Sybron Dental Specialties
Reason for recall
Moisture ingress into packaging of test strips may cause premature failure.
Action taken
Notice was sent to customers of the potential problem on 3/7/2003. The notice included directions to perform a Modified Quality Control Procedure. Instructions are to test strips function each day.
Root cause
Other
Status
Terminated
Product code
JOJ
Quantity affected
4,112 cases of two bottles each. 60 strips per bottle.
Distribution
Nationwide and Canada.
Date initiated
2003-02-25
Date posted
2003-04-09
Date terminated
2004-05-17
Location
Orange, CA