Atlas FDA

Gluteraldehyde Concentration Indicators ''Browne GA Indicator for CIDEX Products''

FDA device recall Z-1480-04 · posted 2004-09-21 · Terminated

Recall details

Recalling firm
Advanced Sterilization Products
Reason for recall
Performance failure complaints.
Action taken
Firm issued a Safety Alert to inform users to perform a daily Quality Control check of the test strips to ensure that they are working.
Root cause
Other
Status
Terminated
Product code
JOJ
Quantity affected
15,334 cases
Distribution
Nationwide, Europe, Latin America, Australia, Canada, and Asia Pacific.
Date initiated
2003-09-04
Date posted
2004-09-21
Date terminated
2005-01-04
Location
Irvine, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
BROWNE GA INDICATOR USE IN 2% ACTIV GLUTARALDEHYDE (K922481)Board of Directors, Albert Browne , Ltd.1994-06-02