Atlas FDA

GLUCm used on Synchron LX and UniCel DxC Synchron Clinical Systems, Part Number: 472500 ( ynchron LX20, LX20 PRO, LXI 72

FDA device recall Z-2388-2010 · posted 2010-09-09 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
UniCel DxC and Synchron LX instruments produced inaccurate (low) GLUCm results. False low results could affect or delay diagnosis or treatment.
Action taken
Beckman Coulter issued an "Urgent: Product Corrective Action" (PCA) letter dated April 2010 to all consignees, identifying the reason for the recall, the affected lots, and the potential for inaccurate GLUCm results due to bubble retention and short samples. Customers were instructed to: 1) establish specific protocols for samples from defined patient populations or sample types, which could include running specimens using GLUCm in duplicate or using cartridge glucose. 2) To reduce the possibility of inaccurate results, ensure that all elements are properly installed and secured, the sample probe is adequately tightened, and proper sample volume and integrity are maintained. A copy of the PCA letter will also be included in all new instrument ship kits. Consignees were instructed to complete and return an enclosed response form. Consignees can contact Beckman Coulter at 1-973-956-5300 .
Root cause
Other
Status
Terminated
Product code
CGA
Quantity affected
84,186 units (79,071 in US & 5,115 in Canada)
Distribution
Worldwide Distribution: USA and Canada
Date initiated
2010-04-15
Date posted
2010-09-09
Date terminated
2013-12-31
Location
Brea, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM (K060256)Beckman Coulter, Inc.2006-02-27
UNICEL DXC 600 AND 800 SYNCHRON SYSTEMS (K042291)Beckman Coulter, Inc.2004-11-12
SYNCHRON LX I 725 CLINICAL SYSTEM (K023049)Beckman Coulter, Inc.2002-10-04
MODIFICATION TO SYNCHRON LX 20 CLINICAL CHEMISTRY SYSTEM (K011213)Beckman Coulter, Inc.2001-05-16
SYNCHRON LX 20 CLINICAL CHEMISTRY SYSTEM, SYNCHRON LX ALBUMIN REAGENT, SYNCHRON LX CREATININE PICRIC REAGENT, SYNCHRON L (K965240)Beckman Instruments, Inc.1997-03-10