Giraffe Incubator infant warmer

FDA device recall Z-0541-2020 · posted 2019-12-03 · Terminated

Recall details

Recalling firm
GE Healthcare, LLC
Reason for recall
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
Action taken
The firm disseminated its notices by letter beginning on 10/11/2019. The notices identified the following safety issues: - The bedside panels of Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation can be upright and look closed but not be latched. - The portholes also can look closed when not latched. - If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. The product may continue to be used following specific safety instructions outlined in the notice. On/about 11/8/2019, the firm issued an "Amended - Urgent Medical Device Recall" letter that supplemented the previous notification to provide additional safety instructions and materials (poster regarding risk of patient fall, user manual addendum, and labeling, including safety labels). Update 6/3/2021: In the event a new consignee was identified, GE Healthcare has sent the original letter.
Root cause
Device Design
Status
Terminated
Product code
FMT
Quantity affected
19425 units
Distribution
worldwide
Date initiated
2019-10-11
Date posted
2019-12-03
Date terminated
2024-08-15
Location
Waukesha, WI