Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked

FDA device recall Z-1022-2013 · posted 2013-03-28 · Terminated

Recall details

Recalling firm
Genetic Testing Institute,inc
Reason for recall
Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an invalid assay is obtained, those results should not be reported and the test should be re-run. An invalid assay would not result in significant harm to a pati
Action taken
Hologic/Gen-Probe sent an Urgent Medical Device Recall letter dated February 4, 2013, via FedEx overnight deliver yto all affected customers. The letter identified affected product, stated the issue, and asked that product from the affected lot stop being used. Reported results for all samples tested using the affected lot should be reviewed. The referring physician should be notified that any negative patient result may be erroneous. Remaining inventory should be returned to Gen-Probe. A response form was also asked to be returned. For questions customers were instructed to contact Technical Support at 262-754-1000 or waukeshatechsupport@gen-probe.com. For questions regarding this recall call 262-754-1026.
Root cause
Process control
Status
Terminated
Product code
LCO
Quantity affected
386 kits (370 US, 16 OUS)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, China, Slovakia, and Spain.
Date initiated
2013-02-04
Date posted
2013-03-28
Date terminated
2016-03-21
Location
Waukesha, WI