GemStar Pump - Battery Product Usage: The GemStar Pump is a small and lightweight, single channel infusion device design
FDA device recall Z-1159-2013 · posted 2013-04-26 · Terminated
Recall details
- Recalling firm
- Hospira Inc.
- Reason for recall
- Battery level lower then 2.4 volts results in corrupt history log and loss of data for preview therapy's parameters and causes an 11/004 error resulting in the pump not being able to be used.
- Action taken
- Hospira sent an Urgent Device Field Correction letter dated March 18, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to notify their consignees if they further distributed the product and confirm they have done so by return of the reply form to Stericycle. Lithium batteries that are older than three (3) years should be replaced. Customers should contact their Hospira Advanced Knowledge Center at 1-800-241-4002, option 4, to determine if their battery need to be replaced and if necessary, arrange for the return of their device to perform battery replacement. Customers with questions should call 1-877-907-7516. For questions regarding this recall call 224-212-4892.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- FRN
- Quantity affected
- List Number 13000: 25,119 pumps; List Number 13100: 185 pumps; List Number 13150: 13,497 pumps
- Distribution
- Worldwide Distribution - USA (nationwide) Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, F
- Date initiated
- 2013-03-18
- Date posted
- 2013-04-26
- Date terminated
- 2017-03-07
- Location
- Lake Forest, IL
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| GEMSTAR SP INFUSION SYSTEM WITH GEMSTAR SP INFUSION SUITE SOFTWARE (K083019) | Hospira, Inc. | 2008-10-22 |
| HOSPIRA GEMSTAR INFUSION PUMP SYSTEM, 7 THERAPY, MODEL 13087; 6 THERAPY, MODEL 13086; PAIN MANAGEMENT, MODEL 13088 (K060806) | Hospira, Inc. | 2006-04-21 |