GEMINI TF Big Bore CT/PET System. The Philips GEMINI PET/CT Imaging Systems are a family of integrated diagnostic X-ray

FDA device recall Z-2199-2015 · posted 2015-07-22 · Terminated

Recall details

Recalling firm
Philips Medical Systems (Cleveland) Inc
Reason for recall
Philips has identified four (4) software defects in the Tumor LOC software application that may potentially result in irradiation of healthy tissue or non-irradiation of diseased tissue.
Action taken
On 7/6/2015 the firm sent Field Safety Notices to their customers.
Root cause
Software design
Status
Terminated
Product code
KPS
Quantity affected
44 Units
Distribution
Worldwide Distribution - US Nationwide, Australia, Belgium, Brazil, Canada, China, Denmark, France, Germany, Israel, Italy, Japan, Lebanon, Netherlands, Portugal, Saudi Arabia, South Africa, Switzerla
Date initiated
2015-07-01
Date posted
2015-07-22
Date terminated
2018-06-26
Location
Cleveland, OH