GEMINI TF 16 PET/CT System & GEMINI TF 64 PET/CT System, Philips Medical Systems, Cleveland, OH The device is a diagnost

FDA device recall Z-0937-2012 · posted 2012-02-02 · Terminated

Recall details

Recalling firm
Philips Medical Systems (Cleveland) Inc
Reason for recall
Philips became aware of an issue relating to a UPS cabinet that contained a melted battery which caused damage to other batteries lower in the cabinet, leading to battery acid leaking on the floor.
Action taken
Philips Medical Systems (Cleveland) Inc communitcated the problem to the facilities via telephone. No customer letter was sent since the correction was already performed by the supplier.
Root cause
Material/Component Contamination
Status
Terminated
Product code
KPS
Quantity affected
2
Distribution
Nationwide Distribution including Maryland and Colorado
Date initiated
2011-10-24
Date posted
2012-02-02
Date terminated
2013-06-18
Location
Cleveland, OH