GEMINI PET/CT Systems Upper Patient Pallet for the GEMINI TF, 16, 64, and 16 Base Slice configurations, Model Numbers: 4

FDA device recall Z-0261-2010 · posted 2009-11-13 · Terminated

Recall details

Recalling firm
Philips Medical Systems (Cleveland) Inc
Reason for recall
Upper patient pallet component of the recalling firm's Gemini PET/CT system may be thinner than normally required which may reduce the product's load bearing capacity, resulting in cracking under less than the designed full-load weight.
Action taken
The firm notified customers with an "URGENT - Field Safety Notice" which was sent to the user/customers via FedEx on 1/26/2009. The letter informs customers of this potential safety hazard; the actions to be taken by the user/customer in order to mitigate risk to patients while undergoing scans; and the actions planned by Philips Medical's Service Engineers in order to correct the problem. The notification refers the user/customers to the applicable sections of their Systems Operations and Systems Information Manuals for a review on procedures for safely loading; unloading; and monitoring patients before, during, and after scanning studies. The notice goes on to inform the customers that replacement parts for the upper patient pallet(s) are on order and that a Philips Healthcare Field Service Engineer will be visiting each customer site within two (2) months in order to replace each affected pallet. Lastly, the notification provides a toll-free telephone number and instructions which the customer can use to contact a local Philips representative if additional information, or assistance is required. The firm's Customer Care Center contact is 1-800-722-9377, option 5: Diagnostic Imaging, option 5: Radiation Therapy, option 3: Oncology.
Root cause
Process change control
Status
Terminated
Product code
JAK
Quantity affected
9 units
Distribution
The recalled products were distributed to customers in the following states within the U. S.: OH, GA, MI, and CA and to customers overseas in: Germany, France Japan, and South Korea.
Date initiated
2009-01-26
Date posted
2009-11-13
Date terminated
2009-11-24
Location
Cleveland, OH

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
GEMINI RAPTOR (K052640)Philips Medical Systems (Cleveland), Inc.2005-10-07