GE Single 20 inch TV Monitor Suspension for XT Bridge, Model 46-240485P6, Catalog Number B2056EG

FDA device recall Z-0853-04 · posted 2004-07-20 · Terminated

Recall details

Recalling firm
General Electric Med Systems
Reason for recall
The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall.
Action taken
Field modification instructions were issued on 3/30/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to hospitals.
Root cause
Other
Status
Terminated
Product code
EAM
Quantity affected
1236 devices
Distribution
The units were distributed nationwide in the United States and worldwide.
Date initiated
2004-03-30
Date posted
2004-07-20
Date terminated
2006-07-24
Location
Waukesha, WI