GE Signa Ovation 0.35 T MR System. The Signa Ovation System is an open, whole body scanner designed to support improved
FDA device recall Z-1556-2009 · posted 2009-06-30 · Terminated
Recall details
- Recalling firm
- GE Medical Systems, LLC
- Reason for recall
- GE Healthcare has identified a potential safety issue with the Signa OpenSpeed and Signa Ovation MR Scanner tables that may impact patient safety. The table may begin lowering on its own while a patient is on the table. If this occurs while the table is in the magnet bore, the table may tilt and could result in patient injury. The patient's head or feet depending on patient orientation, may hit
- Action taken
- GE Healthcare issued an "Urgent Medical Device Correction" letter dated April 16, 2009 addressed to Hospital Administrators, Mangers of Radiology/Cardiology and Radiologists/Cardiologists. The letter describes the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. A GE Healthcare Field Engineer will schedule a field visit to correct the device. Further questions or concerns may be addressed to GE Healthcare at 1-262-521-6681.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- LNH
- Quantity affected
- 203
- Distribution
- Worldwide Distribution -- US including PR and states of AL, AZ, AR, CA, CO,CT, DE, FL, GA, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MN, MS, MO,MT, NV, NH, NE, NJ, NY, NC, OH, OK,OR, PA, SC, SD, TN,
- Date initiated
- 2009-04-16
- Date posted
- 2009-06-30
- Date terminated
- 2013-04-23
- Location
- Waukesha, WI