GE PET Discovery 610, PET Discovery 710. Intended for head and whole body attenuation corrected Positron Emission Tomogr

FDA device recall Z-0024-2014 · posted 2013-10-17 · Terminated

Recall details

Recalling firm
GE Healthcare, LLC
Reason for recall
GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and Screen Save functionality may possibly generate the same Series UID. It has been seen that some PACS may combine series under a patient using only the Series UID
Action taken
Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated March 26, 2013. The letter was addressed to Director of Radiology, Emergency Medicine Physicians and Radiologic Technologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information.
Root cause
Software design (manufacturing process)
Status
Terminated
Product code
KPS
Quantity affected
28 (4 US & 24OUS)
Distribution
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, CO, FL, GA, IL, IA, IS, LA, MA, MO, NH, NM, NY, OH, OK PA, SC, TX, UT, VA, WA, and WI, and the countries of UNITED KINGDOM, DENM
Date initiated
2013-03-26
Date posted
2013-10-17
Date terminated
2013-12-05
Location
Waukesha, WI