GE OEC, Insta Trak 3500 Plus with the Cranial Application for Image Guided Surgery, GE Healthcare, Surgery, Salt Lake Ci
FDA device recall Z-0183-2008 · posted 2008-01-23 · Terminated
Recall details
- Recalling firm
- GE OEC Medical Systems, Inc
- Reason for recall
- Crack, Fracture or Detachment: The Transmitter Arm for the Mayfield-compatible Transmitter may become detached from the unit and enter into the surgical field.
- Action taken
- Consignees were sent an Urgent Recall Notice via mail dated 10/11/07, notifying the customer to examine the plastic end of the transmitter arm and to not use the transmitter arm, if any cracks or damage are observed. Also to stop using Steam Sterilization method and to utilize the other Sterilization methods indicated in the Operators Manual. The root cause is still under investigation GEHC Surgery will provide, at no charge, a solution to this issue when it becomes available.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- LLZ
- Quantity affected
- 169 units
- Distribution
- Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IL, MA, MD, MI, MS, NC, NY, OK, OR, PA, SC, TX, UT, VA, WI, and WV, and countries of Canada, China, France, Hong
- Date initiated
- 2007-10-18
- Date posted
- 2008-01-23
- Date terminated
- 2012-08-10
- Location
- Salt Lake City, UT