Atlas FDA

GE Millennium Systems: MG. Multi-Geometry Digital CSE Dual Detector Gamma Camera. A modular nuclear medicine imaging sys

FDA device recall Z-0414-06 · posted 2006-01-24 · Terminated

Recall details

Recalling firm
GE Healthcare
Reason for recall
In certain cases, one of the four collimator locking mechanisms that attach the collimator to the head may work itself free if collimator exchange has not been performed for an extended period of time. This situation, if neglected, could eventually cause the locking mechanism to unlock itself.
Action taken
Customer letter is being sent to all customers who have an affected system. The letter alerts customers of the potential release of the locking mechanism and requires a Weekly Collimator Exchange Procedure to avoid release of the locking latch until the permanent solution is made available.
Root cause
Other
Status
Terminated
Product code
KPS
Distribution
nationwide including Puerto Rico, and to countries including: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Denmark, Egypt, France, Germany, Iceland, India, Israel, Italy, Jordan, Kor
Date initiated
2005-04-18
Date posted
2006-01-24
Date terminated
2008-02-24
Location
Waukesha, WI