GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, GE Healthcare, Waukesha, WI 53188 (The system consists of an a
FDA device recall Z-0474-2009 · posted 2009-01-21 · Terminated
Recall details
- Recalling firm
- GE Medical Systems, LLC
- Reason for recall
- Software error, computer: GE Healthcare has became aware of an issue with Error Management affecting the Innova 2100, 3100/3100 IQ & 4100/4100IQ. It was reported that when the foot switch was pressed, the live monitor screen was blank and it looked like there were no X-rays being emitted. But the message "Acquisition in progress" was displayed on the DL monitor, even after the foot switch was r
- Action taken
- Consignees sent a "GE Urgent Medical Device Correction" letter dated February 28, 2008. The letter was addressed to Hospital Administrators/Hospital Risk Management Department/Managers of Radiology-Cardiology/ Radiologists-Cardiologists. The letter described the affected products, safety Issue, short term recommendations, long term solution and contact Information. The first batch of letters was received by customers on 5/28/08. For further information, please contact GE Healthcare at 262-544-3894
- Root cause
- Software design
- Status
- Terminated
- Product code
- MQB
- Quantity affected
- 397
- Distribution
- USA including states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL IN, KS, KY, LA, ME, MD, MA, MI,MN, MS,MO, MT, NE, NV, NH, NJ, NM, NY,NC, ND, OH, OK OR, PA, RI, SC, SD, TN, TX UT,VT, VA, WA,
- Date initiated
- 2008-01-30
- Date posted
- 2009-01-21
- Date terminated
- 2016-03-07
- Location
- Waukesha, WI
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| INNOVA 2121IQ, INNOVA 3131IQ, INNOVA 4100, INNOVA 4100IQ, INNOVA 3100, INNOVA 3100IQ, INNOVA 2100IQ WITH INNOVA IVUS OPT (K061163) | Ge Medical Systems, LLC | 2006-06-07 |
| INNOVA 2100-IQ, INNOVA 3100, INNOVA 3100-IQ, INNOVA 4100, AND INNOVA 4100-IQ (K052412) | GE Healthcare | 2005-09-16 |
| INNOVA CT OPTION FOR DIGITAL FLUOROSCOPIC IMAGING SYSTEMS INNOVA MODELS 4100, 4100IQ, 3100, 3100IQ & 2100IQ (K052157) | Ge Medical Systems, LLC | 2005-08-25 |
| INNOVA 4100, INNOVA 3100 (K042053) | Ge Medical Systems, Inc. | 2004-08-26 |
| INNOVA 3100 (K031637) | GE Medical Systems | 2003-08-22 |