GE, Innova 2000 The principle system components include a C-arm, image acquisition, processing and archiving capabilitie
FDA device recall Z-2240-2009 · posted 2009-09-24 · Terminated
Recall details
- Recalling firm
- GE Medical Systems, LLC
- Reason for recall
- GE Healthcare has recently became aware of a potential power supply failure with the Innova 2000, 3100 and 4100 systems that may impact patient safety. A potential power supply failure inside the GE equipment cabinet could occur with no advance warning and thereby prevent x-ray generation and image processing. Should this occur, the system cannot be restarted.
- Action taken
- Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated June 25, 2009. The letter was addressed to Hospital Administrators/Risk Managers, Managers of Radiology/Cardiology or Radiologists/Cardiologists. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. GE Healthcare will correct all affected systems by replacing the power supply located in the GE equipment cabinet. A GE Healthcare service representative will contact you to arrange for this correction. Direct questions about the recall to the firm's Call Center by calling: 1-800-437 -1171, option 4 (United States) or 0120 - 055 - 919 (Japan). For other countries, please contact your local GE Healthcare field service Representative if you have any questions or concerns regarding the notification.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- MQB
- Quantity affected
- 355
- Distribution
- Worldwide Distribution -- US (states of AL, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA., MI, MN, MO, NE, NV, NH, NJ, NY, NC, OH, OK , OR, PA, PR, RI, SC, TN, TX, UT VA, WA,
- Date initiated
- 2009-07-31
- Date posted
- 2009-09-24
- Date terminated
- 2012-05-15
- Location
- Waukesha, WI
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| INNOVA 2000 AND INNOVA 2000S (MOBILE VERSION) (K024200) | Ge Medical Systems, Inc. | 2003-01-16 |